MATERIALS PROVIDED
2. Tramadol test device. The amount of each coated antigen
and/or antibody on the strip is less than 1.0 mg for antigen
conjugate and is less than 1.0 mg for goat anti-mouse IgG
Test zone: contains tramadol bovine protein antigen
Control zone: contains Goat anti-mouse IgG antibody.
3. Conjugate pad: contains mice monoclonal anti-drug
MATERIALS REQUIRED BUT NOT PROVIDED TRAMADOL TEST CASSETTE
FOR THE QUALITATIVE ASSESSMENT OF TRAMADOL
STORAGE AND STABILITY
The test device should be stored at 2 to 30ºC and will be
effective until the expiration date stated on the package. The
product is humidity-sensitive and should be used immediately
after being open. Any improperly sealed product should be
PRECAUTIONS
1. For in vitro diagnostic and forensic use only.
2. Do not use the product beyond the expiration date.
INTENDED USE
3. Handle all specimens as potentially infectious.
4. Humidity sensitive product, do not open foil pouch until it is
immunochromatography based one step in vitro test. It is
designed for qualitative determination of the tramadol and its
5. Use a new urine specimen cup for each sample to avoid
metabolites in human urine specimens at cut-off level of 200
ng/ml. This assay has not been evaluated in the point of care
SPECIMEN COLLECTION AND PREPARATION
location and is for use by Healthcare Professionals only.
It is required that approximately 150 µl of sample for each test.
Fresh urine specimens do not need any special handling or
This assay provides only a preliminary analytical test result. A
treatment. Specimens should be collected in a clean, dry, plastic
more specific alternative chemical method must be used in order
or glass container. If the assay is not performed immediately,
to obtain a confirmed analytical result. Gas chromatography/
urine specimen may be refrigerated at 2-8ºC or frozen up to 7
mass spectrometry (GC/MS) has been established as the
days. Specimens should be thawed and brought to room
preferred confirmatory method by the Substance Abuse Mental
temperature before testing. Urine specimens exhibiting a large
Health Services Administration (SAMHSA). Clinical consideration
amount of precipitate or turbidity should be centrifuged or
and professional judgment should be applied to any drug of
allowed to settle before testing. Avoid contact with skin by
abuse test result, particularly when preliminary positive results
wearing gloves and proper laboratory attire.
QUALITY CONTROL SUMMARY AND EXPLANATION
Good Laboratory Practice recommends the daily use of control
Tramadol is a quasi-narcotic analgesic used in the treatment of
materials to validate the reliability of device. Control materials
moderate to severe pain. It is a synthetic analog of codeine, but
should be assayed as clinical specimen and challenging to the
has a low binding affinity to the mu-opioid receptors. Large
assay cutoff concentration, e.g., 25% above and below cutoff
doses of tramadol can develop tolerance and physiological
concentration. If control values do not fall within established
dependency and lead to its abuse. Tramadol is extensively
range, assay results are invalid. Control materials which are not
metabolized after oral administration. Approximately 30% of the
provided with this test kit are commercially available.
dose is excreted in the urine as unchanged drug, whereas 60%
Tramadol test provides a built-in process control with a different
is excreted as metabolites. The major pathways appear to be N-
antigen/antibody reaction at the control region (C). This control
and O- demethylation, glucoronidation or sulfation in the liver.
line should always appear regardless the presence of drug or
metabolite. If the control line does not appear, the test device
TEST PRINCIPLE
should be discarded and the obtained result is invalid. The
Tramadol test cassette is based on the principle of specific
presence of this control band in the control region serve as 1)
immunochemical reaction between antibodies and antigens to
verification that sufficient volume is added, 2) that proper flow
analyze particular compounds in human urine specimen. The
assay relies on the competition for binding antibody between
drug conjugate and free drug which may be present in the urine
PROCEDURE
specimen being tested. When drug is present in the urine
1. Bring all materials and specimens to room temperature.
specimen, it competes with drug conjugate for the limited
2. Remove the test cassette from the sealed foil pouch.
amount of antibody-dye conjugate. When the amount of drug is
3. Place the transfer pipette in the specimen and depress the
equal or more than the cut-off, 200ng/ml, it will prevent the
binding of drug conjugate to the antibody. Therefore, a positive
4. Hold the pipette in a vertical position over the sample well
urine specimen will not show a colored band on the test line
of the test cassette and deliver 3 drops (120-150 µl) of
zone, indicating a positive result, while the presence of a colored
5. Read the results at 5 minutes after adding the sample.
A control line is present in the test window to work as
procedural control. This colored band should always appear on
INTERPRETATION OF RESULTS
the control line zone if the test device is stored in good condition
Negative:
and the test is performed appropriately.
Two colored bands form. The appearance of two colored bands,
one in test line zone and the other in control line zone, indicates
negative results. The negative result indicates that the tramadol
------------------------------------------------------------------------
concentration in the specimen is either zero or less than cut-off
--------------------------------------------------
Positive:
------------------------------------------------------------------------
One colored band forms. One colored band appears in control
--------------------------------------------------
line zone. No colored band is found in test line zone. This is an
indication that the tramadol level in the specimen is above the
Invalid:
------------------------------------------------------------------------
If there is no colored band in control line zone, the test result is
---------------------------------------------------
invalid. Retest the sample with a new device.
Each listed substance that commonly found in the urine was
evaluated and indicated negative result at concentration of 100
Note: A borderline (+/-) in test line zone should be considered LIMITATION OF PROCEDURE
The assay is designed for use with human urine only. A positive
result with any of the tests indicates only the presence of a
drug/metabolite and does not indicate or measure intoxication.
There is a possibility that technical or procedural error as well
other substances in certain foods and medicines may interfere
with the test and cause false results. Please refer
“SPECIFICITY” section for lists of substances that will produce
either positive results, or that do not interfere with test
Phencyclidine Phenylethylamine-α Phenylpropanolamine
performance. If a drug/metabolite is found present in the urine
11-nor-∆8 –THC-9-COOH 11-nor-∆8 –THC-9-COOH
specimen, the assay does not indicate frequency of drug use or
distinguish between drug of abuse and certain foods and
EXPECTED RESULTS
Immunospec Tramadol Test is a qualitative assay. It identifies
REFERENCES
tramadol in human urine at a concentration of 200 ng/ml or
1. Urine testing for drugs of abuse, NIDA Research
higher. The concentration of the tramadol cannot be determined
by this assay. The test is intended to distinguish negative result
2. Drug Facts and Comparisons, 55th ed. St. Louis: Facts
from presumptive positive result. All positive results must be
confirmed using an alternate method, preferably GC/MS.
3. Raffa, R. et al. Mechanism of action study , J. of
Pharmacol. Exp. Ther. Vol. 267, 331 ( 1993)
PERRFOMANCE CHARACTERISTICS A. Sensitivity
The cut-off concentration (sensitivity level) of tramadol test is
European Authorized Representative:
CEpartner4U , Esdoornlaan 13, 3951DB Maarn
. The Netherlands. Tel.: +31 (0)6.516.536.26
B. Precision
The precision study was performed by three individuals
observing the test result to determine the random error of visual
interpretation. The test results were found to have no significant
differences between the three observers.
Manufacturer: C. Specificity
The specificity for tramadol test was tested by adding various
drugs, drug metabolites, and other compounds that are likely to
be present in urine. All compounds were prepared in drug-free
1. Interference testing
Tramadol test performance at cut-off level is not affected when
pH and Specific Gravity ranges of urine specimen are at 4.5 to
The following substances were tested and confirmed not to
interfere with tramadol test at the listed concentrations.
2. Specificity
The following table lists compounds that are detected by
tramadol test which produced positive results when tested at
levels equal or greater than the concentrations listed below:
Case study tumour type: 51 year old male with melanoma Case study tumour type: Melanoma Medical history prior to first course of radiowave therapy (RWT) prior to 3.07.2006 Presenting symptoms Patient is a 51 year old male who presented to his GP in 2004 with a mass in the left scapular area. Initial diagnosis – February 2004 – Melanoma Method of diagnosis – Hi
Dolphin ® 9500/9550 Mobile Computer The Dolphin 9500 series mobile computers deliverunparalleled performance and value for all mobiledata collection applications. Their lightweight,ergonomic, and rugged designs provide the ease ofuse and operational reliability needed to excel inrobust mobile applications such as package delivery,field service, route accounting, transportation